Formaldehyde
— EPA TO - 11A
EPA Method TO-11A is used for sampling of formaldehyde and
other carbonyl compounds (aldehydes and ketones) in ambient
air using a coated-solid adsorbent. Refer to the method for
further details (http://www.epa.gov/ttn/amtic/files/ambient/airtox/to-11ar.pdf).
Equipment:
- Air sampling pump capable of sampling at the recommended
flow rate with the sampling medium in-line
- Air flow calibrator.
- SKC Sorbent Tube Cat. No. 226-120
Sampling Guide:
Sampling Flow Rate - 100 to 1500 ml/min
Sample Time – varies with volume required.
Air Volume – varies (50 to 100 L for indoor air samples;
see Table).
| Sampling
Rate
(L/min) |
Formaldehyde Reporting
Limit (ppb) |
| Sampling Duration (min)
|
| 30 |
60 |
120 |
240 |
480 |
| 0.2 |
4.8 |
2.4 |
1.2 |
0.59 |
0.30 |
| 0.5 |
1.9 |
0.95 |
0.48 |
0.24 |
0.12 |
| 1.2 |
0.79 |
0.40 |
0.20 |
0.10 |
0.05 |
Sampling and Analysis:
- Break open both ends of a sorbent tube.
- Insert the sorbent tube into the rubber sleeve of the
adjustable low flow holder or tube holder. The arrow on
the sorbent tube indicates air flow and should point toward
the pump (the end of the tube with the YELLOW section should
point towards the pump).
- Calibrate the pump using a separate sorbent tube than
the one used for sampling. The pump should be calibrated
to the flow rate specified for the method.
- To collect a sample, use a new sorbent tube and set it
up as described above. Do not break off the ends until ready
to sample. The section of the tube with the WHITE sorbent
should be facing away from the pump.
- Set up tube in the location where you wish to collect
the sample. Sampling tube should be in a vertical position.
Turn the pump on and note starting time.
- Sample at known flow rate for the recommended period
of time to achieve the desired reporting limit (Table).
Check sampling pump from time to time to ensure that it
is operating properly.
- At the end of the sampling period, turn the pump off
and note the time. Remove the sorbent tube, seal the ends
with the red caps provided, label the tube and record any
pertinent sampling information.
- Calibrate the pump with the representative sampling media
in line to verify that the flow has not changed by more
than 5%.
- Submit field blanks from the same lot number as the sample
tubes. Field blanks should be subjected to exactly the same
handling as the samples (open, seal and transport) except
that no air is drawn through them.
- Carefully pack sample tubes and field blanks in a cooler
with FROZEN ice packs. Be sure to include all pertinent
information (e.g. sample identification, sampling date,
time and sample volume, etc.,) on chain of custody form
that is submitted with the samples. Secure cooler and then
ship it to the laboratory via overnight courier.
Storage and Shipping:
Store in freezer before use. After sampling, refrigerated
storage and shipping to the laboratory is required.
Analyzing Method:
High-performance liquid chromatography - Ultraviolet detector
(HPLCUV)
|