Welcome to Alpha Environmental  
 
 

Formaldehyde — EPA TO - 11A

EPA Method TO-11A is used for sampling of formaldehyde and other carbonyl compounds (aldehydes and ketones) in ambient air using a coated-solid adsorbent. Refer to the method for further details (http://www.epa.gov/ttn/amtic/files/ambient/airtox/to-11ar.pdf).

Equipment:

  1. Air sampling pump capable of sampling at the recommended flow rate with the sampling medium in-line
  2. Air flow calibrator.
  3. SKC Sorbent Tube Cat. No. 226-120

Sampling Guide:
Sampling Flow Rate - 100 to 1500 ml/min
Sample Time – varies with volume required.
Air Volume – varies (50 to 100 L for indoor air samples; see Table).

Sampling
Rate
(L/min)
Formaldehyde Reporting Limit (ppb)
Sampling Duration (min)
30
60
120
240
480
0.2
4.8
2.4
1.2
0.59
0.30
0.5
1.9
0.95
0.48
0.24
0.12
1.2
0.79
0.40
0.20
0.10
0.05

Sampling and Analysis:

  1. Break open both ends of a sorbent tube.
  2. Insert the sorbent tube into the rubber sleeve of the adjustable low flow holder or tube holder. The arrow on the sorbent tube indicates air flow and should point toward the pump (the end of the tube with the YELLOW section should point towards the pump).
  3. Calibrate the pump using a separate sorbent tube than the one used for sampling. The pump should be calibrated to the flow rate specified for the method.
  4. To collect a sample, use a new sorbent tube and set it up as described above. Do not break off the ends until ready to sample. The section of the tube with the WHITE sorbent should be facing away from the pump.
  5. Set up tube in the location where you wish to collect the sample. Sampling tube should be in a vertical position. Turn the pump on and note starting time.
  6. Sample at known flow rate for the recommended period of time to achieve the desired reporting limit (Table). Check sampling pump from time to time to ensure that it is operating properly.
  7. At the end of the sampling period, turn the pump off and note the time. Remove the sorbent tube, seal the ends with the red caps provided, label the tube and record any pertinent sampling information.
  8. Calibrate the pump with the representative sampling media in line to verify that the flow has not changed by more than 5%.
  9. Submit field blanks from the same lot number as the sample tubes. Field blanks should be subjected to exactly the same handling as the samples (open, seal and transport) except that no air is drawn through them.
  10. Carefully pack sample tubes and field blanks in a cooler with FROZEN ice packs. Be sure to include all pertinent information (e.g. sample identification, sampling date, time and sample volume, etc.,) on chain of custody form that is submitted with the samples. Secure cooler and then ship it to the laboratory via overnight courier.

Storage and Shipping:
Store in freezer before use. After sampling, refrigerated storage and shipping to the laboratory is required.

Analyzing Method:
High-performance liquid chromatography - Ultraviolet detector (HPLCUV)



Alpha Environmental, Inc. 11 Ackerman Ave. Emerson, NJ 07630

Telephone: 201-420-0600

Email Us

Copyright World Wide 2004 - All Rights Reserved - Alpha Environmental, Inc.